The question
What is conductivity actually measuring, and what can and cannot be concluded from it?
What the measurement is
Conductivity measures how readily water conducts an electric current, which depends on the concentration of dissolved ions in it. More dissolved salts, more conductivity.
That makes it a practical stand-in for total dissolved solids, and therefore the basis on which cycles of concentration and blowdown are controlled. It is fast, continuous, cheap to measure and robust enough for a plant room, which is why it is the parameter most cooling systems are actually run on.
What it cannot tell you
Conductivity is a sum, not a breakdown. It responds to total ionic content and says nothing about which ions are present.
- It cannot distinguish scaling species from corrosive ones. The same reading could be mostly calcium and alkalinity, or mostly chloride — two very different problems.
- It does not measure inhibitor residual. Treatment chemical contributes to conductivity, but the reading is not a check on the programme.
- It does not see suspended solids or biological load, neither of which is ionic.
- It does not indicate scaling tendency on its own — that requires the specific ion balance, temperature and pH.
Getting a reading you can trust
Conductivity varies significantly with temperature, so a compensated reading is essential — and the compensation has to be set for the water being measured, not left at a default that assumes something else.
Sensor type follows the duty. Contacting cells are accurate in clean water at lower ranges but are affected by fouling and coating. Electrodeless sensors tolerate fouling and higher conductivity, which is why they suit dirtier recirculating systems.
Location decides whether the number means anything. The sensor needs to sit in flowing, well-mixed water — not in a dead leg, and not immediately downstream of the chemical injection point, where it will read a slug of concentrate rather than the system.
Using it for control
Blowdown driven by conductivity holds cycles of concentration at a measured value. When evaporation raises the concentration above setpoint, the valve opens; when concentration falls back into the deadband, it closes.
The deadband is not a detail. A setpoint without one produces a valve that cycles constantly, wearing the actuator and destabilising the reading. Feed limits and lockout timers serve the same purpose — they are what make an automatic system stable rather than merely automatic.
The most useful thing conductivity gives you is not the instantaneous number but the trend. A slow upward drift points to blowdown not keeping up or a valve partially blocked. A sudden fall suggests dilution — a leak, an overflow, or make-up entering somewhere it should not.
What to check
- 01Confirm temperature compensation is enabled and correctly configured.
- 02Calibrate the sensor against a known standard and record the pre-calibration reading.
- 03Check the sensor is in flowing, representative water — not a dead leg or a post-injection point.
- 04Verify the sensor type suits the water: electrodeless where fouling is likely.
- 05Compare the controller reading against a handheld meter periodically.
- 06Confirm the blowdown setpoint has a sensible deadband.
- 07Trend the reading rather than reading it — the shape carries more information than the value.
When to get technical support
If the controller reading and a calibrated handheld disagree, resolve that before making any control decision — a drifting sensor makes every dosing and blowdown decision downstream of it wrong, quietly, for as long as it goes unnoticed. And if conductivity is stable but deposits or corrosion are appearing, the answer is not in this measurement: a full water analysis is needed to see which ions are actually present.
Talk to an EngineerFrequently asked
No. It gives you the total, not the composition. Two waters with identical conductivity can have completely different scaling and corrosion behaviour depending on which ions make up that total. Conductivity is for control between analyses, not instead of them.